Precise Results The METTLER TOLEDO HRT (HiRes Technology) includes two built-in weights. Together with FACT (Fully Automatic Calibration Technology), this allows the balance to automatically perform adjustment and linearization. |
![]() |
Efficient Handling Reduce the number of movements needed for every weighing. Open the draft shield ErgoDoor with one hand and simultaneously place the weighing sample with your other hand. |
![]() |
Space Saving Save working space thanks to split draft shield doors that do not project beyond the balance, and sideways mounted connectors at the back of the balance. |
![]() |
Easy Cleaning Cleaning has never been this easy: dismantle the QuickLock draft shield in a few short steps, then wash all glass panels and the drip tray in the dishwasher. |
![]() |
Standard Features |
All models offer | |||||||||||||||||||
|
|
Limit values | |
Maximum capacity | 220 g |
Readability | 0.1 mg |
Repeatability (at nominal load) 1) | 0.09 mg (200 g) |
Repeatability (at low load) 1) | 0.07 mg (20g) |
Linearity deviation | 0.2 mg |
Sensitivity offset 2) | 0.8 mg (200 g) |
Typical values | |
Repeatability (at nominal load) 1) | 0.07 mg (200 g) |
Repeatability (at low load) 1) | 0.05 mg (20g) |
Minimum weight (according to USP) | 150 mg |
Minimum weight (U=1%, 2 sd) | 10 mg |
Settling time | 2 s |
Sensitivity offset 2) | 0.6 mg (200 g) |
Dimensions | |
Weighing pan dimensions (mm) | 80 |
Balance dimensions LxWxH (mm) | 358x247x331 |
1) Standard deviation (sd) 2) Stability of sensitivity with FACT self-adjustment switched on |